Glucose Aqueous Solution Sensing by Modified Hilbert Shaped Microwave Sensor

This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Submitted: 2026-07-23
CC BY-NC 4.0 This work is licensed under Creative Commons Attribution–NonCommercial International License (CC BY-NC 4.0).

Abstract

We present a microwave sensor based on the modified first order Hilbert curve designed for detecting glucose concentration in aqueous solutions by using a real-time microwave near-field electromagnetic interaction technique. We observed S11 reflection parameters of the sensor at resonant frequencies depend on the glucose concentration. We could determine the glucose concentration in the 0-250 mg/dl concentration range at an operating frequency near 2.6 GHz. The measured minimum detectable signal was 0.018 dB/(mg/dl) and the measured minimum detectable concentration was 1.68 mg/dl. The simulation result for the minimum detectable signal and the minimum detectable concentration was 0.023 dB/(mg/dl) and 1.3 mg/dl, respectively. Our proposed system has excellent potential to serve as a human bloodless glucometer.